Chengdu Ultrapure Technology Co.,LTD Logo
Customer Service Hotline:86-400-884-6567、86-028-87573647
    • ULZGUP-TEZ countertop Lab Water Purification Systems

    • Publisher:公司官网 Publication Date:2018-01-09 13:56:03 Page Views:9828
    • 优普UPL落地式(超)纯水机

       

       

      ULUP®-TEZ Countertop (Ultra) Water Purifier
            The ULZGUP-TEZ countertop (ultra) water purifier produces water that meets the requirements of GB/T6682-2008, GB/T 33087-2016 and other standards by taking urban tap water that meets the hygiene standards (GB5749-2022) for drinking water as raw water, and adopting a multi-stage water treatment process. It is an efficient laboratory high-end (ultra) water purifier that is of all-plastic ABS engineering plastic housing, resisting acid, alkali and corrosion, with a 4.3'' touch screen display, presenting friendly and intuitive human-computer dialog.System

      Functions:

      1.PLC HMI system, easy to operate, optional English interface, intuitive display;

      2. Real-time display of process flow, enabling mastery of operating status;

      3. Dual-stage RO + modularized EDI design, effectively improving water quality and reducing operation and maintenance cost;

      4. Multi level permission management, multi-user login permissions, preventing illegal operations and avoiding data conflicts;

      5. Monitoring of four-way water quality, detection and display of multi-way flow, accurate, intuitive and reliable;

      6. Reliable historical data recording and USB download, enabling easy tracing and fault retrieval for over 5,000 (scalable) pieces of data;

      7. Qualitative, quantitative and quick intake, optional intake handle, assisting in long-distance intake;

      8. Intelligent multi-way circulation system, combined with excessive discharge function, achieving simultaneous sterilization and bacteria inhibition;

      9. Unique patented double-layer PE sterile water tank is adopted, equipped with a respirator, safe and pollution-free for water storage, more convenient for cleaning;

      10. Efficient dual wavelength ultraviolet digester, effective sterilization and digestion to reduce TOC, optional TOC online display function, meeting the needs of high-end users;

      11. Intelligent management of consumables, easier for maintenance.

      Technical Parameters

      Product Series

      ULZGUP-TEZ Series

      Model

      ULZGUP-I-10/20TEZ

      ULZGUP-II-10/20TEZ ULZGUP-III-10/20TEZ ULZGUP-IV-10/20TEZ

      Inlet Water

      TDS: <200ppm, water pressure: 0.10~0.40MPa, water temperature: 5~45°C

      Water production rate

      10/20 (L/h)

      UV

      N/A

      60W  185/254nm 60W  185/254nm

      UF Membrane

      N/A

      5000Dalton N/A 5000Dalton
      Type2 Water Quality

      Resistivity≥10MΩ.cm @25℃(real-time monitoring)

      Type3Water Quality

      Conductivity:1~5μS/cm(real-time monitoring)

      Type1Water Quality Resistivity:18.2MΩ.cm @25℃(real-time monitoring

      Microparticle

      ≤1 pc./ml

      TOC

      ≤15ppb ≤10ppb ≤8ppb ≤5ppb

      微生物

      ≤1CFU/ml
      Rnases <0.5pg/ml
      Dnases <5pg/ml

      Heavy Metal Ion

      ≤0.1ppb

      Pyrogen

      N/A ≤0.001EU/ml N/A ≤0.001EU/ml

      Terminal Micro-filtration

      0.22μm imported terminal microfilter

      Dimension of Principal Machine

      380Wx 505D x 620H ( mm )

      Weight of Principal Machine

      Net weight: approx. 35 (kg)

      Working Power Supply

      AC220V~50HZ (Power: 200W~410W)

      Water Tank Configuration

      30L ULUPURE sterile water tank (double-layer medical-grade PE, standard respirator and faucet, optional UV)

      Scope of Application

      AAS, AES, IC, HPLC, ICP/MS, VOC, IVF, PCR, liquid/gas chromatography, amino acid analysis, protein purification, electrophoresis/gel analysis, toxicological analysis, animal and plant cell culture, molecular biology, gene study, test-tube baby......

      NoteDimension: W (Front) × D (Side) × H (Height), with an error of 0-5mm.

      Ordering Instructions

      1. The water yield of the product refers to the rated yield of RO membrane when the water temperature is 25 ℃. For every 1 ℃ decrease in water temperature, the water yield of the RO membrane decreases by about 3%. When the water temperature is below 5 ℃, the RO membrane will stop producing water. When selecting and ordering, users must consider the water temperature drop in winter in their region.

      2. Optional English operation system, TOC display module, pretreatment module, water leakage protection module, UPHD-I-B water intake module.
      3. The conductivity is a typical value for continuous water production.
      4. The TOC indicator of the produced water is related to the TOC value of the raw water and the system configuration.
      5. The water quality entering the system will directly affect the quality of the water produced and the lifespan of the filter cartridge.
      6. In areas with poor raw water quality (TDS≥200ppm), a fully automatic water softener is recommended to effectively reduce the consumption of consumables for subsequent systems.